https://diabetes.acponline.org/archives/2013/07/12/9.htm

Abbot FreeStyle Insulinx meters recalled

The Abbott FreeStyle Insulinx blood glucose meters were recently recalled because at extremely high blood glucose levels of 1,024 mg/dL and above, the meters will display and store in memory an incorrect test result that is 1,024 mg/dL below the measured result.


The Abbott FreeStyle Insulinx blood glucose meters were recently recalled because at extremely high blood glucose levels of 1,024 mg/dL and above, the meters will display and store in memory an incorrect test result that is 1,024 mg/dL below the measured result.